Bench Boost by Inorganic Ventures

ICP Considerations for Lithium
Send us Fan Mail
This week on Bench Boost our team reviews special considerations for measuring lithium by ICP-OES and ICP-MS. They discuss common lithium sample types and preparation guidance. Autumn emphasizes the impact of high total dissolved solids (TDS) and easily ionizable elements (EIEs) causing plasma ionization-balance shifts and signal suppression for some sample types. Additional tips include selecting appropriate internal standards, choosing suitable emission lines, and avoiding error propagation during large dilutions.
Audio and video play straight from the show's own hosting. PodSubs stores and serves no media.